General introduction:

  • ppbRAE 3000 is a series of single-target gas detection machines, used to measure VOC organic gas

Go to the product page, see more parameters VOC ppbRAE 3000 gas detector

Summary of instructions for use:

  • How to open the machine: Press and hold the power button on the ppbRAE 3000 gas detector until the screen displays the RAE logo system, the power button will start the software itself, check the profile configuration itself, and then go into normal measurement operation when the screen displays as follows:
Voc ppbRAE 3000 Organic Gas Detection Manual
Voc ppbRAE 3000 Organic Gas Detection Manual
  • Calibration and zero machine: Press and hold the power and N/- buttons simultaneously until the screen requires a password, if entering a pass, the program will take us to install software including: Calibration, gas type and unit of measurement, alarm, data storage, and finally the machine configuration. If you do not enter the pass but want to go always into the calibration process, press the Select button to enter the machine calibration including clean air calibration (zero) and calibration with gas bottle (span)
  • Machine settings: Press the power and N/- buttons simultaneously until the screen requires a pass, the default pass of the machine when the factory is 0000, press Y/+ to enter the pass then press the power button to confirm, then the screen will go to the menus to enter the machine settings, including
    1. Calibration: Optional Zero calibration and Span calibration
    2. Gas type and unit of measurement setting is ppb or ppm
    3. Alarm settings including: High alarm, low alarm, STEL and TWA, alarm type, siren light
    4. Data saving settings: Data deletion options, save type, save area, save time period…
    5. Operation mode settings: Allows the installation of profiles for the machine to operate by default for the next time.

Some charts describing machine software:

Calibration software chart and zero ppbRAE 3000
Calibration software chart and zero ppbRAE 3000
PpbRAE 3000 parameters settings menu chart
PpbRAE 3000 parameters settings menu chart
Table explaining the settings for the VOC gas detector of the RAE system
Table explaining the settings for the VOC gas detector of the RAE system

Purchase: There are 2 codes to choose from as follows

by TES Industry.vn

Related Post

Leave a Reply