The MiniRAE 3000 gas detector offers the most advanced organic gas (VOC) monitoring solution on the market whether in the field of industrial hygiene, rust detection or toxic material handling. With sensor technology (PID) – photoionization detector, this VOC gas detector can measure gas concentrations in a wide range of 0~15,000 ppm, with a response time of only 3 seconds. The machine has more than 200 parameters pre-installed to adjust the sensor accuracy when measuring different gases, in addition, it is equipped with PID sensor self-cleaning technology that RAE system has patented. All of which says that the MiniRAE 3000 can measure a lot of chemical gases with faster speed and accuracy than any machine on the market.
It is equipped with a good power sampling pump, which can draw gas both horizontally and vertically with a distance of up to 100 ft. The machine can optionally upgrade the configuration with Bluetooth connectivity, in this way this VOC gas detector can transmit measurement results, real-time alarms at a distance of up to 2 miles when equipped with a raelink3 wireless router and a command center. The machine is equipped with a flashlight for convenient measurement in low light conditions, and is equipped with a high cover with waterproof, dustproof dick – which ensures that it always works well in the most extreme conditions.
MiniRAE 3000 PGM-7320 Gas Meter Features:
- Bringing the most advanced features to the handheld VOC gas detector on the market
- PID sensor technology offers a wide range of 0-15,000 ppm
- Reply to results within 3 seconds
- Self-compensate for humidity
- Patented PID Lamp Self-Cleaning Technology
- Option to add a Bluetooth communication connection
- Data can be forwarded at a distance of up to 2 miles and integrated suction pumps with the ability to suck up to 100 ft
- Preset more than 200 precision self-adjusting parameters
- Built-in flashlight
- IP65/IP67 Dust and Waterproof Cover
- Record data for up to 6 months with a sampling distance of 1 minute
- Variety of 10 languages to choose from
Application:
- Emergency Rescue
- Environmental Cleanup / Pollution Treatment
- Treatment of toxic materials
- Gas leak detection
- Urban search and rescue
- Protect the scene
- Toxic exposed worker areas
STANDARD EQUIPMENT INCLUDES: (May change at time of order)
- MiniRAE 3000 Monitor, Model PGM-7320
- Datalogging with ProRAE Studio Package for Windows 95, 98, 2000, NT, ME &XP
- Charging/download adapter
- RAE 10.6eV lamp, 9.8eV or 11.7eV optional if specified
- Flex-I-Probe
- External filter
- Protective rubber boot
- Alkaline battery adapter
- Lamp cleaning kit
- Tool Kit
- Rechargeable Lithium-ion (Li-ion) battery
- Operation & Maintenance Manual & CD-ROM
- Soft leather carrying case
TECHNICAL SPECIFICATION
Gas detection | VOCs |
Range (ppm) | 0 to 999.9 |
Resolution (ppm) | 0.1 |
Response Time (T90) | < 3 sec |
Gas 2 | VOCs |
Range (ppm) | 1000 to 15,000 |
Resolution (ppm) | 1 |
Response Time (T90) | < 3 sec |
Size | 10″ L x 3.0″ W x 2.5″ H (25.5 cm x 7.6 cm x 6.4 cm) |
Weight | 26 oz (738 g) |
Sensors | Photoionization sensor with standard 10.6 eV lamp |
Battery | Alkaline battery adapterExternal field-replaceable lithium-ion battery pack |
Operating Time | 16 hours of operation (12 hours with alkaline battery) |
Display Graphic | 4 lines, 28 x 43 mm, with LED backlight for enhanced display readability |
Keypad | 1 operation and 2 programming keys, 1 flashlight on/off |
Direct Readout | VOCs as ppm by volume / High values / STEL and TWA / Battery and shutdown voltage / Date, time, temperature |
Alarms | 95dB buzzer (at 12″/30 cm) and flashing red LED to indicate exceeded preset limits / High: 3 beeps and flashes per second / Low: 2 beeps and flashes per second / STEL and TWA: 1 beep and flash per second / Alarms latching with manual override or automatic reset / Additional diagnostic alarm and display message for low battery and pump stall |
EMI/RFI | Highly resistant to EMI/RFI. Compliant with EMC directive (2004/108/EC); R & TTE directive (1999/5/EC) |
IP Rating | IP-65 (unit running): protected against dust, protected against low-pressure jets of water from all directions / IP-67 (unit off and without flexible probe): protected against dust, protected against immersion between 15cm and 1m depth. |
Data Logging | Standard 6 months at one-minute intervals |
Calibration | 2-point or 3-point calibration for zero and span. Calibration memory for 8 calibration gases, alarm limits, span values and calibration dates |
Sampling Pump | Internal, integrated flow rate at 500 cc/minSample from 100′ (30m) horizontally and vertically |
Low Flow Alarm | Auto shut-off pump at low-flow condition |
Communication | Download data and upload instrument setup from PC via charging cradle |
Certifications | US and Canada: Classified as Intrinsically Safe for use in Class I, Division 1 Groups A, B, C, D/Europe: ATEX II 2G EEx ia IIC T4 |
Operating Temperature | -4° to 122° F (-20° to 50° C) |
Humidity | 0 to 95% relative humidity (non-condensing) |
Attachments | Durable bright yellow rubber boot |
See more product information on the company’s website here
Actual images:
VOC Gas Meter Backup Battery Box VOC Meter PID Lamp Cleaner MiniRAE 3000 Computer Connection Cable MiniRAE 3000’s Monitor only kit MiniRAE 3000 Documentation Backup filter for MiniRAE 3000 VOC MiniRAE 3000 Air Meter Travel Charger Tools included with VOC MiniRAE 3000 VOC MiniRAE 3000 Gas Meter Display Annual maintenance service of VOC gas detectors VOC MiniRAE 3000 Gas Meter